- Question
- Is the ACURATE neo2 valve non-inferior to the Evolut platform in patients with symptomatic severe aortic stenosis and selected anatomies undergoing TAVI?
- Design
- Investigator-initiated, open-label, randomised, multicentre, non-inferiority trial with a 2x2 factorial design, conducted at 10 sites in Germany with 835 randomised patients.
- Population
- Patients with symptomatic severe aortic stenosis and selected anatomies undergoing transcatheter aortic valve implantation.
- Intervention
- TAVI with the self-expanding ACURATE neo2 valve.
- Control
- TAVI with an Evolut valve (PRO, PRO+ or FX).
- Primary outcome
- 30-day composite of all-cause mortality, stroke, moderate/severe prosthetic valve regurgitation and permanent pacemaker implantation. Non-inferiority was tested in the intention-to-treat population with an absolute margin of -6%.
- Results
- The primary endpoint occurred in 60/389 patients (15.4%) with ACURATE neo2 and 120/395 (30.4%) with Evolut (rate difference 15.0%; lower boundary of the 1-sided 95% CI 10.1%; p for non-inferiority <0.001; 2-sided 95% CI 9.1-20.7; p for difference <0.001). Permanent pacemaker was implanted in 42/374 (11.2%) vs 101/381 (26.5%) (p <0.001).
- Safety
- Moderate/severe prosthetic valve regurgitation was low in both groups (1.3% vs 1.7%; p=0.6).
- Conclusion
- In selected anatomies, TAVI with ACURATE neo2 was non-inferior to Evolut, with a lower primary endpoint rate driven by fewer pacemaker implantations at 30 days.
- Why it matters
- It showed that the specific design of the ACURATE neo2 valve, although withdrawn from the market, can reduce the need for pacemakers and may inform the design of future TAVI platforms.