- Question
- Does routine use of a cerebral embolic protection device during transcatheter aortic-valve implantation (TAVI) reduce the incidence of periprocedural stroke?
- Design
- Randomized, controlled trial (1:1) across 33 centers in the United Kingdom, with 7635 participants assigned to TAVI with or without a cerebral embolic protection device.
- Population
- Patients with aortic stenosis undergoing TAVI.
- Intervention
- TAVI with a cerebral embolic protection device (CEP group), 3815 participants.
- Control
- TAVI without a cerebral embolic protection device (control group), 3820 participants.
- Primary outcome
- Stroke within 72 hours after TAVI or before hospital discharge, if discharge occurred sooner.
- Results
- The primary outcome occurred in 81 of 3795 participants (2.1%) in the CEP group and in 82 of 3799 (2.2%) in the control group (difference, -0.02 percentage points; 95% CI, -0.68 to 0.63; P = 0.94). Disabling stroke occurred in 47 participants (1.2%) with CEP and 53 (1.4%) in the control group.
- Safety
- Death occurred in 0.8% of the CEP group and 0.7% of controls. Access-site complications were similar (8.1% vs. 7.7%). There were 24 serious adverse events in 22 participants (0.6%) with CEP and 13 in 13 participants (0.3%) in the control group.
- Conclusion
- Among patients undergoing TAVI, routine use of cerebral embolic protection did not decrease the incidence of stroke within 72 hours.
- Why it matters
- The trial challenges the routine use of these devices in all patients undergoing TAVI, as no clinical benefit in preventing periprocedural stroke was shown.